OverviewSpecificationsResources
Process analyzers for the determination of dissolved nickel
The WATERLINE Ni10 process photometer can be used for the monitoring and control of the dissolved nickel concentration in water. Measurement principle is the photometric determination of nickel by forming a red complex after the reaction with Dimethylglyoxim. Main application for the photometer is the monitoring of the nickel concentration in process water.
The analyzer consists of a control unit with touchscreen and an analysis unit with measuring chamber, valve, dosing pumps and all required tube connections. The control unit includes a microprocessor which controls the automatic measurement incl. sampling, rinsing, reagent dosing and surveillance of the photodetection system.
Advantages:- Automatic measurement incl. self test and drift compensation
- Easy operation via touchscreen
- Adjustable limit value and alarm value
- Programmable analog output (0/4-20 mA), optionally with USB port for easy data storage
- Adjustable break time between two analysis
- External start/stop of an analysis possible
- No external calibration required
- External plug connections (IP65) for alarm relay, limit relay, analysis relay, external start/stop, analog output 0/4-20 mA
- Multi range power supply (110–230 Volt, 50–60 Hz)
- Including polycarbonate wall cabinet
Parameter | Nickel (dissolved) |
Description | Automatic microprocessor controlled analyzer for the photometric determination of dissolved nickel |
Typical Applications | Control of the nickel concentration in process water |
Analysis Method: | Photometric determination of dissolved nickel by the Dimethylglyoxim method |
Analyzer type | WATERLINE Ni10 |
Measuring Range | 0,02 – 1,00 ppm |
Resolution | 0,01 ppm |
Accuracy | 3 % of end value |
Reproducibility | 2 % of end value |
Zero-point Stability | automatic adjustment |
Number of Samples | 1 |
Sample | |
---|---|
Operating Pressure | 0,1 - 10 bar |
Temperature | 5 - 30 °C |
Sample Volume | 25 ml per analysis (excluding rinsing) |
Sample Condition | clear, filtrated |
Chemical Demands | pH 4 – 9, Fe2+ < 0,01 mg/l, Cu2+ < 0,01 mg/l |
Drain | pressure free into open drain |
Reagents | |
Number | 3 |
Storage Temp. | 5 – 20 °C |
Usage/analysis | appr. 0,27 ml / 0,54 ml / 0,54 ml |
Reagent volume | 250 ml / 500 ml / 500 ml |
Suitable for | appr. 925 analysis |
Analysis | |
Cycle (approx.) | 7 min (incl. rinsing time) |
Sample interval | 1 – 99 min or external start/stop |