OverviewSpecificationsResources
The Right Features To Measure, Analyze And Control pH And Temperature
- Accurate and stable readings.
- 5 programmable on/off relays: 4 for pH and 1 for temperature.
- Manual or automatic temperature compensated (ATC) pH readings.
- Programmable hysteresis for customized control.
- Isolated and reversible 4-20 mA current output for pH. This output can be used as a proportional output.
- DC voltage output for pH/ORP pre-amplifier.
- RS-485 output for remote access via an IBM compatible computer. This allows multiple units to be connected, ganged, together.
Range | |
---|---|
pH | - 2.00 to 16.00 pH |
Temp | - 10.0 to 120.0 °C |
Resolution | |
pH | 0.01 pH |
Temp | 0.1 °C |
Accuracy | |
pH | ± 0.01 pH ± 1 LSD |
Temp | ± 0.1 °C ± 1 LSD |
General | |
pH: | |
pH Temp Compensation | Auto -10.0 to 120.0 °C |
pH Buffer Recognition | 4.00, 4.01, 6.86, 7.00, 9.18, 10.01 |
pH Buffer Temp. Range | 0 to 60 °C |
PH Electrode Offset Recognition | ± 100 mV at pH 7.00 |
- 91.7 to +108.3 mV at pH 6.86 | |
>pH Electrode Slope Recognition | ± 30 % at pH 4.00, 4.01, 9.18 or 10.01 |
Input Impedance | >1013 ohms |
Calibration End Point Sensing | Yes |
Pre-amplifier Output | ± 5 V DC |
Temperature: | |
Temperature Sensor | Thermistor, 10.00 k ohm at 25 °C |
4-20 mA Analog Output: | |
Current Output Range | 4 to 20 mA (isolated), programmable span |
Current Output Scale | User programmable, linear or log |
Maximum Load | 500 ohm |
Accuracy | ± 0.02 mA |
Isolation Voltage | 500 V DC |
Relays: | |
Control Type | 5 ON/OFF controls |
Relay Output | 5 A at 115 V AC or 2.5 A at 220 V AC, resistive load only |
Other: | |
Power Supply | 115 V AC or 230 V AC at 50/60 HZ |
Security | 4-digit password |
Audio Feedback | All Touch Keys |
Communication | RS-485 |
Fuse | 0.315 A / 250 V fast acting glass tube |
Display | 128 x 64 graphical LCD with backlight |
Ambient Temp. Range | 0 to 50 °C |
Case | IP65, 1/4 DIN case, 148 mm depth |
Weight | 950 g |