Toxicity analyzer

A Product by Orlab
OverviewSpecificationsResources

Industry-leading toxicity detection


The Microtox® Model 500 (M500) analyser is a laboratory-based, temperature-controlled, self-calibrating photometer that measures acute toxicity. Microtox® M500 is a biosensor-based measurement system that uses bioluminescence technology to monitor for either accidental or deliberate contamination of both water supplies and waste water. With over 2,400 instruments sold worldwide, the Microtox® toxicity test system is the industry standard for rapid toxicity screening and analysis. DeltaTox® II is the portable toxicity analyser used with the Microtox® technology

Microtox® rapid toxicity detection is an in vitro test system that uses bioluminescent bacteria for the detection of toxicity in water and is used as a screening system to detect the relative toxicity of a sample. Results are given either as an effective concentration (EC) value or as a % inhibition. Applications include the testing of samples containing biological toxins, industrial effluent, industrial process waters, municipal effluent, drinking water eco toxicological samples, hazardous waste, soil, sediments, storm water and medical products for bio reactivity.

The Microtox® M500 has a built-in cooling block, enabling it to be used for ISO and DIN standard methods. The Omni software contains detailed step by step protocols for 11 standard methods, including ISO 11348-3, DIN 38412 and ASTM D5660.

  • Photometer pre-programmed for COD of all ranges (LR, HR and HR+)
  • Filters: 420 and 620nm
  • Cell holder: 16mm round cells
  • Power requirements: 230V 50/60Hz AC as well as with rechargeable battery
Size 18.3cm x 39.1cm x 41.2cm (7” x 15” x 16”)
Weight 9.5kg (21lbs)
Power 100 +/- 10% VAC, 2 Amp Slo Blo, 50/60 Hz
120 +/- 10% VAC, 2 Amp Slo Blo, 50/60 Hz
220 +/- 10% VAC, 1 Amp Slo Blo, 50/60 Hz
240 +/- 10% VAC, 1 Amp Slo Blo, 50/60 Hz
Room Temperature Requirement 15ºC to 30ºC
Temperature REAGENT well 5.5ºC ± 1ºC
Acute Mode
Incubator block 15ºC ± 0.5ºC
READ well 15ºC ± 1.0ºC
Mutatox/Chronic Mode
Incubator block 27ºC ± 0.5ºC
READ well 27ºC ± 1.0ºC
It is available in 120V and 220V versions
Resources