Amperometric type Chlorine Dioxide Analyser
- Range: 0.05 to 10.00 ppm or 0.05 to 20.00 ppm
- Automatic temperature compensation by an integrated temperature sensor.
- No consumables & no reagents are required for analysis.
- Controller housed in ABS Plastic (Ip65)
- Built-In relays for alarms.
- 16X2 Alphanumeric LCD display
- Potable water
- Swimming pools
- Drinking water
- Process water
- Service water
Measuring Ranges : | 0.05 to 10.00 ppm or 0.05 to 20.00 ppm | |
Resolution : | 0.05 | |
Measuring Principle : | Membrane Covered, amperometric 2- electrode system | |
Sensor working temperature : | > 50 to < 450 C | |
Sample Flow Rate : | Approx. 30 L/h | |
Sample pH range : | pH 1 to pH 12 | |
Sample : | 0.5 bar | |
Response Time : | 90=Approx. 2 min | |
Slope Drift : | Approx. <1% per month | |
Sample Temperature compensation : | Automatically, by an integrated temperature sensor | |
Relays : | Two potential – free contacts, programmable | |
Display : | 16X2 Alphanumeric LCD display | |
Slope Calibration : | By using pocket colorimeter, DPD-1 Method | |
Ambient Temperature : | 00C to 550C | |
Relative Humidity : | 10-99% at 250C | |
Power Supply : | 230v AC (24v DC optional) | |
Power Consumption : | 10VA | |
Auxiliary Power Supply : | 24v DC @ 30mA | |
Communication : | RS232 | |
Retransmission Output : | 4-20mA | |
Mounting : | Wall Mounting | |
Measuring Transmitter enclosure : | ABS Plastic (IP 65) | |
Dimensions : | 130X185X105mm | |
Process Connection : | Sample Inlet : | Push Fit 8mm OD |
Sample Outlet : | Push Fit 8mm OD | |
Weight : | Approx. 5 Kg | |
Ordering Information : | ||
Waterline 1000 Series : | ||
Cat No. OR-CDA10 Range : | 0.05-10 ppm | |
Cat No. OR-CDA20 Range : | 0.05-20 ppm |