The OCMA-500 Oil Content Monitor has been redesigned for even better operability, maintaining its user-friendly features.After injecting the sample, all you have to do is press a button to get the monitoring operation done quickly, from oil extraction to sample measurement and draining.
Flexible and powerful Features of OCMA-500 Oil Content Monitor
- Fully automated one-touch operation, from oil extraction to sample measurement and draining
- Reduced environmental impact and running costs (20% cut in solvent consumption)
- With no need for a solvent evaporation process, the OCMA-500 offers easy operation, and can detect even oils with low boiling points.
- Color graphic LCD
- USB flash memory drive port
- Backlit extraction tank
- Unit Conversion
Measurement principle: | Solvent extraction, NDIRanalysis (Infrared spectroscopy) |
Measurement range: | 0mg/l ~ 200mg/l dynamic ranging |
Resolution: | 0 to 99.9 mg/L; 0.1 mg/L 100 to 200 mg/L; 1 mg/L |
Repeatability: | 0 to 9.9 mg/L; ± 0.2 mg/L ± 1 digit 10.0 to 99.9 mg/L; ± 2 mg/L ± 1 digit 100 to 200 mg/L; ± 4 mg/L ± 1 digit |
Display Method: | 3.5 inches Backlight Colour graphic LCD |
Calibration Method: | Zero, Span Calibration |
Extraction Solvent: | S-316 solvent |
Amount of test sample required: | Sample water : Solvent = 2 : 1 |
Amount of extraction solvent required: | 8 mL |
Extraction Method: | Built-in extractor |
Ambient Operationg Temperature: | 0 to 40 ? (non condensation) |
Power supply: | AC 100V to 240V ±10%, 50/60 Hz |
External Dimensions: | 200(W)×313(D)×342(H) mm |
External Output: | Output to an USB memory stick |