Process analyzers for the determination of Chlorine
The WATERLINE CL2 and WATERLINE CHLOR process photometers can be used for the monitoring and control of the chlorine concentration in water. Measurement principle is the photometric determination of chlorine by dosing a special reagent based on N,N-Diethyl-pphenylendiaminsulfat (DPD) to a buffered water sample. The wine-red colour formed as a result of the reaction is detected by a monochromatic photometric detection system. The WATERLINE CL2 determines the concentration of free chlorine or of total chlorine. The WATERLINE CHLOR determines the concentration of free chlorine and of total/bounded chlorine. The analyzers consist of a control unit with touchscreen and an analysis unit with measuring chamber, valve, dosing pumps and all required tube connections. The control unit includes a microprocessor which controls the automatic measurement incl. sampling, rinsing, reagent dosing and surveillance of the photodetection system. Main applications for the photometers are the monitoring of the chlorine concentration in drinking water, cooling water and swimming pool water.
- Automatic measurement incl. self test and drift compensation
- Easy operation via touchscreen
- Adjustable limit value and alarm value
- Programmable analog output (0/4-20 mA, AQUACON CHLOR has 2 analog outputs)
- Optional: USB port for easy data storage
- Optional: data transfer via wireless network
- Adjustable break time between two analysis
- External start/stop of an analysis possible
- No external calibration required
- External plug connections (IP65) for alarm relay, limit relay, analysis relay, external start/stop, analog output 0/4-20 mA
- Multi range power supply (110–230 Volt, 50–60 Hz)
- Including polycarbonate wall cabinet
Parameter | Chlorine (free and or total/bounded) | |
Description | Automatic microprocessor controlled analyzer for the photometric determination of Chlorine | |
Typical Applications | Control of chlorination plants (drinking water, pools, cooling water) | |
Analysis Method: | Photometric determination of Chlorine (DPD method) | |
Analyzer type | WATERLINE CL2 | WATERLINE CHLOR |
Measuring Range | 0,02 – 2,00 ppm Cl2 (free or total) | 0,02 – 2,00 ppm Cl2 (free) 0,02 – 2,00 ppm Cl2 (total or bounded) |
Resolution | 0,01 ppm | |
Accuracy | 2 % of end value | |
Reproducibility | 1 % of end value | |
Zero-point Stability | automatic adjustment | |
Number of Samples | 1 | |
Sample | ||
---|---|---|
Operating Pressure | 0,1 - 10 bar | |
Temperature | 5 - 30 °C | |
Sample Volume | 25 ml per analysis (excluding rinsing) | |
Sample Condition | clear, filtrated | |
Chemical Demands | pH 4-8 | |
Drain | pressure free into open drain | |
Reagents | ||
Number | 2 | 3 |
Storage Temp. | 5 – 25°C | 5 – 25°C |
Usage/analysis | appr. 0,27 ml / each reagent | appr. 0,27 ml / each reagent |
Reagent volume | 250 ml / 250 ml | 250 ml / 250 ml / 250 ml |
Suitable for | appr. 925 analysis | appr. 925 analysis |
Analysis | ||
Cycle (approx.) | 3 - 5 min (excluding flushing time) | |
Sample interval | 1 – 99 min or external start/stop |