Waterline F10/F20/F30 (Fluoride)

OverviewSpecificationsResources

Process analyzers for Fluoride measurement

The WATERLINE F10/F20/F30 process titrators can be used for the monitoring and control of the fluoride concentration in water. Measurement principle is the determination of dissolved fluoride with a special developed ion-sensitive electrode (ISE). Therefore, two reagents (TISAB and standard solution) are added to the water sample in the measurement chamber of the analyzer (standard addition method with titration). Main applications for the ion-sensitive analyzers are the monitoring of the fluoride concentration in process water and drinking water.

The analyzers consist of a control unit with touchscreen and an analysis unit with measuring chamber, valve, dosing pumps and all required tube connections. The control unit includes a microprocessor which controls the automatic measurement incl. sampling, rinsing, reagent dosing and surveillance of the ion-sensitive electrodes.

Advantages:
  • Automatic measurement incl. self test and drift compensation
  • 3 different measurement ranges available
  • Easy operation via touchscreen
  • Adjustable limit value and alarm value
  • Programmable analog output (0/4-20 mA)
  • Adjustable break time between two analysis
  • External start/stop of an analysis possible
  • No external calibration required
  • External plug connections (IP65) for alarm relay, limit relay, analysis relais, external start/stop, analog output 0/4-20 mA
  • Multi range power supply (110–230 Volt, 50–60 Hz)
  • Including 2-part polycarbonate wall cabinet
Parameter Fluoride
Description Microprocessor-controlled analyzer for the determination of Fluoride in water
Typical Applications Monitoring and control of process water and drinking water
Analysis Method: Fluoride determination with ion-sensitive electrode
Analyzer type WATERLINE F10 WATERLINE F20 WATERLINE F30
Measuring Range 0,1 – 10,0 ppm 1 – 100 ppm 5 – 500 ppm
Resolution 0,01 ppm 1 ppm 1 ppm
Accuracy 5 % of end value
Reproducibility 3 % of end value
Zero-point Stability automatic adjustment
Number of Samples 1
Sample
Operating Pressure0,1 - 10 bar
Temperature5 - 30 °C
Sample Volume100 ml per analysis (excluding rinsing)
Sample Conditionclear, with particles < 0.5 g/l ; < 50 µm
Chemical DemandspH 4-8
Drainpressure free into open drain
Reagents
Number2
Storage Temp.5 – 30°C
Usage/analysisStandard: concentration dependend
TISAB: appr. 2 ml/analysis
Reagent volume500 / 500 ml
Suitable forappr. 250 analysis
Analysis
Cycle (approx.)15 min., incl. rinsing time
Sample interval1 – 99 min or external start/stop